1 question

Calibration

Three intervals under 29 CFR 1910.95(h)(5): a functional check before each day's use, per (h)(5)(i); an acoustic calibration at least annually, per (h)(5)(ii) and Appendix E, or sooner if the daily check deviates by 10 dB or more; and an exhaustive calibration at least every two years, per (h)(5)(iii) or sooner if the annual acoustic check deviates by 15 dB or more.

audiometer calibration intervals, and which gets recorded