FAQ 1 min read Calibration Equipment Selection

How often do audiometers need to be calibrated?

Three intervals under 29 CFR 1910.95(h)(5): a functional check before each day's use, per (h)(5)(i); an acoustic calibration at least annually, per (h)(5)(ii) and Appendix E, or sooner if the daily check deviates by 10 dB or more; and an exhaustive calibration at least every two years, per (h)(5)(iii) or sooner if the annual acoustic check deviates by 15 dB or more.

Share
LinkedIn X

Three intervals under 29 CFR 1910.95(h)(5): a functional check before each day's use, per (h)(5)(i); an acoustic calibration at least annually, per (h)(5)(ii) and Appendix E, or sooner if the daily check deviates by 10 dB or more; and an exhaustive calibration at least every two years, per (h)(5)(iii) or sooner if the annual acoustic check deviates by 15 dB or more.

OSHA requires the daily check be performed, not recorded — (h)(5)(i) says the audiometer “shall be checked before each day’s use.” What the audiometric record must carry, under (m)(2)(ii), is the date of the last acoustic or exhaustive calibration. Worth closing the gap anyway: an undocumented daily check is indistinguishable from one that never happened if an audiogram is ever challenged. Note also what (h)(5)(i) asks for — testing a person with known, stable thresholds, and listening to the output for distortion. It is a human-in-the-loop check by design, which is worth confirming when a system advertises an automated one. Systems that log the check as part of the testing workflow close the evidence gap structurally, not by reminder.

Found this useful? Share it.
LinkedIn X

Two ways to start. Both take 15 minutes.

See PureTest in action, or get pricing built for your program.